Remove the foil cover from the cuvette. The liquid surface should lie between the two lines marked on the cuvette. Do not touch the clear fl at surfaces in the lower part of the cuvette.
For more detailed information, please consult the instructions for use
Remove the foil cover from the cuvette. The liquid surface should lie between the two lines marked on the cuvette. Do not touch the clear fl at surfaces in the lower part of the cuvette.
Fill the capillary with the sample (20 µl) up to the white stopper. Ensure that there are no air bubbles in the capillary. Wipe away any residual sample from the outside of the capillary.
Place the sample (20 µl) into the buffer solution in the cuvette and dispense by pressing down the plunger. Make sure that the capillary is completely empty.
Close the cuvette tightly with a CRP reagent cap. Do not press down the blue inner part of the cap. Once the sample is added to the buffer, it may be held for up to two hours prior to running the assay.
Choose Measure on the display of the QuikRead go instrument and insert the cuvette in the measurement well of the instrument. The bar code should be facing you. The display shows how the measurement is proceeding.
When the measurement is completed, the result is shown in the display and the cuvette will rise up from the measurement well automatically.
The QuikRead go® instrument is a fully automatic small size photometer used for quantitative or qualitative measurements with QuikRead go reagent kits.
The QuikRead 101 is a compact, portable, multi-analyte POC system used globally throughout the health-care sector in more than 40 countries.